MΩ/cm
3
0 to 19.99 0.01
uS/cm
3
0 to 2.000 0.01
0 to 20.00 0.1
0 to 200.0 0.1
0 to 2000 1.0
mS/cm
3
0 to 20.00 10
0 to 200.0 50
Section I - Specifications Page 2
MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL
Section I - Specifications
pH ORP Conductivity Flow
Display
Front Panel: 4 x 7 segment 1/2” LED display, 1 LED indicator 0n-line,7 LED Bar Graph
Inside Panel: 2 x 16 alpha-numeric LCD display
Power Requirements 120Vac (±10%) 50/60Hz (less than 12VA) or 240Vac (±10%) 50/60Hz (less than 12VA)
Flow: 0 to 9999 with selectable
flow rate units
Volume: 0 to 9999 with Auto Range
Measuring Range pH: 0.01 to 14.00 ORP: -1999 to +1999mV
(Dependent on sensor)
Flow rate units: Gallons (GP), Cubic
Temp: 0 to 100°C or Temp: 0 to 100°C or
Feet (CF), Liters (LP), Cubic Meters
32° to +212°F 32° to +212°F (
CM), custom by entering factor
related to Gallons
Time units: Seconds (S),
Temp: 0 to 100°C or 32° to +212°F
Minutes (M), Hours (H)
Automatic or Manual
Temperature Automatic or Manual
Not required
User selectable temperature
Not required
Compensation 0 to 100°C (32° to +212°F)
compensation slope 0.0 to 10.0%/°C.
0 to 100°C (32° to +212°F)
Temperature Unit °C or °F Not required
Temperature Sensor User selectable: 300Ω NTC Thermistor, 3000Ω NTC Thermistor or Pt. 1000 RTD Not required
Auto-Calibration
Manual Calibration
Dry Calibration
Calibration Modes Manual Calibration
Temperature Calibration
Sample Calibration K factor Input
Temperature Calibration Temperature Calibration
Ambient Conditions Temperature: -20°C to +60°C or -4°F to +140°F Humidity: 0 to 90% RH (non-condensing)
Menu Access
Auto-Calibration, Manual
Manual-Calibration, Manual Calibration
Front Panel
Calibration,Temperature
Temperature Display Temperature Display
Not available
Display
Menu Access Inside Panel Full Access to all parameters of operations menu
Sensor to SHARK Differential Sensor: 3000 ft
300 ft 2000 ft
Distance Combination Sensor: 10 ft
Two Control Relays, 10A / NO, 5A / NC @ 240VAC or 28VDC.
Mode: Process control, Adjustable parameters: process direction,(rising or falling) on-set-point, off set-point,
(0 to 100% of full scale), cycle timer (on / off, 0 to 600 seconds), failsafe (on / off).
Relay Outputs
One Alarm Relay, 10A / NO, 5A / NC @ 240VAC or 28VDC.
Mode: High / Low Alarm, Adjustable parameters: Low on / Low off set-point (0 to 100% of full scale, low on must be less than
low off), High On / High Off set-point (0 to 100% of full scale, High on must be greater than High off).
4 to 20mA Channel 1
Isolated Output, Range expand 0 - 100% of full scale (min segment 10% of full scale), max. load 800Ω
Analog Outputs 4 to 20mA Channel 2
Isolated Output, Range expand 0 - 100% of full scale (min segment 10% of full scale), max. load 800Ω
Can be set to track temperature if sensor is equipped with a temperature sensor
Memory Back-up All user settings are retained indefinitely in memory (EEPROM)
Mechanical
Enclosure: NEMA 4X, 1/4 DIN, polycarbonate enclosure with four 1/2” conduit holes
Mounting: Universal Mounting kit for surface, pipe and panel mount, is included
Sensor Input
Probe: -600 to +600mV Probe: -1999 to +1999mV Cell: 0 to 9999Ω
Paddle: 0 to 2000Hz
Temp. Sensor: 0 to 9999Ω Temp. Sensor: 0 to 9999Ω Temp. Sensor: 0 to 9999Ω
Invalid Entries Invalid entries cannot be stored
Manual Test Mode Process value can be simulated with arrow keys to verify correct setup of outputs
Manual Relay Override Relays can be set to on / off / auto, to verify correct wiring of auxiliary devices, or to manually adjust process
Output Hold All outputs are placed on hold when SHARK is in Menu mode
Recall data from last
Recall data from last calibration, calibration mode,
calibration, calibration buffer
Calibration Data
1st & 2nd accepted buffer value and probe mV output,
accepted value, and cell Recall store K factor.
calibration temperature, calibration slope, and probe
resistance, calibration
efficiency
temperature
Auto Return User selectable auto return if SHARK is left in menu mode or if relays are left in manual override mode for more than 10 min.
Display Damping User can select rate at which SHARK updates display. Enables display damping of unstable process
Net Weight 2.2lbs (1kg)
Approvals ULC (pending)